A Novel <sup>141</sup>Ce-Based Flood Field Phantom: Assessment of Suitability for Daily Uniformity Testing in a Clinical Nuclear Medicine Department.
Author | |
---|---|
Abstract |
:
Daily quality control testing of a γ-camera is of the utmost importance in assessing whether the camera is suitable for clinical use. The aim of our study was to assess the suitability of a fillable Ce-based flood field phantom developed in-house for daily quality control testing of γ-cameras. Daily uniformity testing was performed for 113 d using the fillable Ce phantom and a commercially available sheet-type Co phantom, and the results were compared. The average integral uniformity obtained by the Ce and Co phantoms was 3.24% and 2.72%, respectively, for detector 1 and 3.31% and 2.78%, respectively, for detector 2. The Ce phantom we developed is a suitable alternative to the commercially available Co phantom. |
Year of Publication |
:
2017
|
Journal |
:
Journal of nuclear medicine technology
|
Volume |
:
45
|
Issue |
:
3
|
Number of Pages |
:
225-229
|
ISSN Number |
:
0091-4916
|
URL |
:
http://tech.snmjournals.org/cgi/pmidlookup?view=long&pmid=28408699
|
DOI |
:
10.2967/jnmt.116.189340
|
Short Title |
:
J Nucl Med Technol
|
Download citation |